Analyst (Lond), 2014, 139(11):2867-2872

Design of molecular beacons: 3’ couple quenchers improve fluorogenic properties of a probe in real-time PCR assay

Convenient preparation of fluorogenic hairpin DNA probes (molecular beacons) carrying a pair of FAM fluorophores (located close to 5’-terminus of the probe) or a pair of BHQ1 quenchers on 3’-terminus (with (BHQ1)2or BHQ1–BHQ1 composition) is reported. These probes were used for the first time in a real-time PCR assay and showed considerable improvements in fluorogenic properties (the total fluorescence increase or signal-to-background ratio) in assay conditions vs. conventional one-FAM-one-BHQ1 molecular beacon probes as well as vs. hydrolyzable one-FAM-one-BHQ1 TaqMan probes. At the same time, such multiple modifications of the probe do not influence its Cq (a fractional PCR cycle used for quantification). The probe MB14 containing a BHQ1–BHQ1 pair showed a PCR fluorescence/background value of 9.6 which is more than two times higher than that of a regular probe MB2 (4.6). This study demonstrates prospects for the design of highly fluorogenic molecular beacon probes suitable for quantitative real-time PCR and for other potential applications (e.g. intracellular RNA detection and SNP/mutation analysis). © 2014 the Partner Organisations.

Ryazantsev DY, Kvach MV, Tsybulsky DA, Prokhorenko IA, Stepanova IA, Martynenko YV, Gontarev SV, Shmanai VV, Zavriev SK, Korshun VA

IBCH: 4444
Ссылка на статью в журнале: http://xlink.rsc.org/?DOI=C4AN00081A
Кол-во цитирований на 11.2023: 17